8
Using the Stellar device for the first time
When using the Stellar device for the first time, ensure that all components are in working
condition and perform a functional test (see “Performing a functional test” on page 13). Alarms
should also be checked by the physician when setting up a new patient (see Clinical Guide).
Working with other optional accessories
Attaching the pulse oximeter
Use of a pulse oximeter may be recommended by your clinician.
Contraindication
The pulse oximeter does not meet defibrillation-proof requirement per IEC 60601-1: 1990,
clause 17.h.
1 Connect the plug of the finger pulse sensor to the plug of the pulse oximeter.
2 Connect the plug of the pulse oximeter at the rear of the device.
To view the oximetry values, from the Monitoring menu, select Monitoring.
Adding supplemental oxygen
Oxygen may be prescribed by your clinician.
Note: Up to 30 L/min can be added.
WARNING
Oxygen flow must be turned off when the device is not operating, so that unused
oxygen does not accumulate within the device and create a risk of fire.
Use only certified, clean oxygen sources.
ResMed strongly recommends adding oxygen into Stellars oxygen inlet at the
rear of the device. Entraining oxygen elsewhere, ie into the breathing system via
a side port or at the mask, has potential to impair triggering and accuracy of
therapy/monitoring and alarms (eg, High Leak alarm, Non-vented mask alarm). If
used in this way, therapy and alarm operation must be verified each time oxygen
flow is adjusted.
Oxygen supports combustion. Oxygen must not be used while smoking or in the
presence of an open flame. Only use oxygen supply in well-ventilated rooms.
The breathing system and the oxygen source must be kept at a minimum
distance of 2 m away from any sources of ignition (eg, electrical devices).
Oxygen must not be used while the device is being operated within the mobility bag.
1
2

Other manuals for ResMed stellar series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the ResMed stellar series and is the answer not in the manual?

ResMed stellar series Specifications

General IconGeneral
Data StorageSD card
Device TypeVentilator
Modes of VentilationCPAP, S, S/T, T
Power SupplyAC 100-240V, 50/60Hz
Operating Temperature5°C to 40°C
Operating Humidity10% to 95% (non-condensing)
Operating AltitudeUp to 2500 meters
Sound Level30 dBA
HumidificationIntegrated heated humidifier
ConnectivityWi-Fi

Related product manuals