RESPIRONICS*
Chapter
2:
Warnings,
Cautions,
&
Notes
Operational
(Continued)
A
physician
must
prescribe
use
of
a
humidifier
with
the
Millennium.
Use
of
only
bubble
type
humidifi-
ers
is
recommended
and
connections
must
be
secure
and
free
of
leaks.
Federal
(U.S.)
law
restricts
this
device
to
sale
by,
or
on
the
order
of,
a
licensed
physician.
This
device
should
be
used
only
under
the
supervision
of
a
physician.
If
the
Millennium
Oxygen
Concentrator
has
been
subjected
to
sub
freezing
temperatures
for
an
ex-
tended
period
of
time,
it
should
be
allowed
to
warm.
up
to
the
stated
operating
temperatures
before
power
up.
Failure
to
do
so
could
result
in
improper
performance
and
or
alarm
conditions
until
the
unit
reaches
normal
operating
temperatures.
Do
not
turn
on
unit
while
any
filter
is
wet
or
moist.
2.2
CAUTIONS
s
Place
the
unit
where
cooling
airflow
is
unrestricted.
Be
sure
to
inform
the
user
not
to
position
the
unit
close
to
drapery
or
curtains
that
might
restrict
the
airflow.
*
When
using
liquid
leak
detector,
be
careful
not
to
allow
it
to
contact
electrical
parts.
+
Make
sure
connections
of
fittings,
tubing,
and
hoses
are
secure.
*
Be
cautious
when
using
thread
sealants
because
they
can
cause
extensive
damage
to
the
internal
parts
of
the
unit
if
allowed
within
tubing
or
fittings.
e
Clean
all
exterior
cabinet surfaces
periodically
by
wiping
with
a
damp
cloth,
using
a
mild
detergent
and/or
hospital
disinfectant.
+
Use
only
Respironics
or
factory-authorized
replace-
ment
parts
and
accessories.
1002403
Millennium
Oxygen
Concentrator
System
Service
Manual
2-4