iBreeze CPAP Clinical Guide
21
Alert
Type
Possible Cause
Action
The respiration tube
expired, please replace.
Notification
In standby state, the
usage time exceed the
respiration tube setting
service life.
Click “Confirm”,
replace the respiration
tube.
The water tank expired,
please replace.
Notification
In standby state, the
usage time exceed the
water tank setting service
life.
Click “Confirm”,
replace the water tank.
The filter expired,
please replace .
Notification
In standby state, the
usage time exceed the
filter setting service life.
Click “Confirm”,
replace the filter
The mask expired,
please replace.
Notification
In standby state, the
usage time exceed the
mask setting service life.
Click “Confirm”,
replace the mask
It is time for device
maintenance ,Please
contact the service
provider for device
maintenance.
Notification
In standby state, the
usage time exceed the
setting device
maintenance time.
Click “Confirm”,
contact the service
provider for device
maintenance.
Low Minute Volume
(MV).
Alert 0
Minute volume less than
the setting threshold
value.
Check the respiration
tube or adjust therapy
parameter setting.
High Respiratory Rate
(RR).
Alert 0
Respiratory Rate exceed
the setting threshold
value.
Check the respiration
tube or adjust therapy
parameter setting.
Low Respiratory Rate
(RR).
Alert 0
None.
Check the respiration
tube or adjust therapy
parameter setting.
Apnea.
Alert 1
An apnea is detected and
exceeds the setting
duration.
Check the respiration
tube or adjust therapy
parameter setting.
High Leakage Volume.
Alert 1
1. Inappropriate
connection of mask and
respiration tube.
2. The water tank is not
plugged in.
Check the connection
of the mask or
respiration tube, and
the connection of the
water tank.
High Inspiration
Pressure.
Alert 0
During breath cycle
monitored pressure is
higher than the set
pressure threshold.
Check the respiration
tube or adjust therapy
parameter setting.
Low Expiratory
Pressure.
Alert 0
During breath cycle
monitored pressure is
lower than the set
pressure threshold.
Check the respiration
tube or adjust therapy
parameter setting.
Mask vent holes blocked.
Alert 1
Mask vent holes is blocked.
Check the mask.
Low Tidal Volume (Vt).
Alert 0
During breath cycle,
Average Vt less than
0.6*Setting Value.
Check the respiration
tube or adjust therapy
parameter setting.