iBreeze CPAP Clinical Guide
20
8 Device Alert
There are 4 types of alerts described here:
Alert 0: Gray bottom white tips, no light, no sound, always display, disappear until the
prompt condition not met.
Alert 1: Orange bottom white tips, blue-ray flashing, with sound, always display, disappear
until the prompt condition not met.
Alert 2: Red bottom white tips, blue-ray flashing, with sound, exclusive alarm until the user
opens the alert to confirm or press the Therapy On/Off key.
Notification: White reminder message, no light, no sound, after the corresponding process,
the message automatically disappears.
Alert Summary Table: The following table summarizes the alerts.
1. Pressure sensor failure,
flow sensor failure, blower
failure in therapy state; 2.
Power board short-circuit
3. Humidifier heating
circuit short-circuit