CPAP Clinical Guide
iBreeze 20C / 20C Pro
iBreeze 20A / 20A Pro

Questions and Answers

  • A
    Anna GibbsAug 18, 2025
    How to fix air leaking from around my Resvent iBreeze 20A Medical Equipment mask?
    • K
      Karen SilvaAug 18, 2025
      If air is leaking from around your mask, it might not be fitted correctly. Ensure your mask is properly fitted. Refer to your mask user guide for fitting instructions or use the Mask Fit function to check the fit and seal.
  • L
    Laurie WelchAug 21, 2025
    What to do if my Resvent iBreeze 20A Medical Equipment device doesn't turn on?
    • L
      Lauren LittleAug 21, 2025
      If your device isn't turning on, check the outlet to ensure it has power and that the device is properly plugged in. Verify the AC power cord is correctly connected to the power supply and the power supply cord is securely connected to the device’s power inlet. If the problem persists, contact your home care provider, returning both the device and power supply for assessment.
  • A
    alexismillerAug 23, 2025
    Why is my mouth so dry using Resvent iBreeze 20A Medical Equipment?
    • T
      Trevor BaileyAug 23, 2025
      If your mouth is very dry and uncomfortable while using the device, it could be due to air escaping through your mouth. Try increasing the Humidity Level. Alternatively, consider using a chin strap to keep your mouth closed or switch to a full face mask.
  • C
    Charles FoxAug 26, 2025
    How to stop a dry nose with Resvent iBreeze 20A Medical Equipment?
    • S
      Sandra CarlsonAug 26, 2025
      If you're experiencing a dry or blocked nose, the humidity level might be set too low. Adjust the Humidity Level to a more comfortable setting.
  • D
    Deanna BeltranAug 28, 2025
    What to do if I get water droplets with Resvent iBreeze 20A Medical Equipment?
    • J
      Justin OchoaAug 28, 2025
      If you're getting droplets of water on your nose, in the mask, and air tubing, the humidity level might be set too high. Adjust the Humidity Level to reduce condensation.
  • D
    Daniel SmithAug 30, 2025
    Why is the air pressure too low in my Resvent Medical Equipment mask?
    • D
      Deborah FletcherAug 30, 2025
      If the air pressure in your mask seems too low, the Ramp function might be in progress. Either wait for the air pressure to build up or turn the Ramp Time off.
  • S
    Steven PittmanSep 2, 2025
    Why is my Resvent Medical Equipment water tank leaking?
    • C
      Caleb MahoneySep 2, 2025
      If your water tank is leaking, it may not be assembled correctly, or it could be damaged or cracked. Check for any damage and reassemble the water tank correctly. If the leak persists, contact your care provider for a replacement.
  • A
    Angela RobertsSep 4, 2025
    Why does my Resvent Medical Equipment keep blowing air after stopping therapy?
    • S
      Steven RomeroSep 4, 2025
      The device blows a small amount of air after therapy has stopped to prevent condensation in the air tubing. It will stop automatically after a few minutes.
  • D
    David ClaySep 7, 2025
    What to do if Resvent Medical Equipment display is erratic?
    • D
      Dustin RoySep 7, 2025
      If the device's display is erratic (crashes, blank or blue screen), it may be due to the device being dropped or mishandled. Unplug the device, then reapply power. If the problem continues, contact your home care provider.
  • L
    Linda WallaceSep 10, 2025
    What to do for a non-responsive key on Resvent iBreeze 20A?
    • D
      davisdouglasSep 10, 2025
      If you are experiencing a key exception (non-responsive or insensitive), it may be due to program crashes or key misalignment. Unplug the device, then reapply power. If the problem continues, contact your home care provider.

Summary

3 Contraindications

5 System Overview

6 Operation

6.1 Accessing Clinical Mode

Guides on how to access, view, and set therapy and device configuration parameters.

6.5 Clinical System Settings

6.6 Clinical Alert Settings

7 Therapy Information

7.1 CPAP Mode

Explains the CPAP mode, delivering constant pressure throughout therapy.

7.2 CPAP-V Mode

Describes CPAP-V mode, a titration mode combining CPAP and APAP.

7.3 APAP Mode

Details APAP mode, where pressure adapts automatically based on patient status.

7.4 APAPW Mode

Specifies APAPW mode for female patients with enhanced sensitivity.

7.5 Events

Explains how the device detects and responds to sleep events like apnea and hypopnea.

Automatically Pressure Regulation Rules

Details the rules governing automatic pressure adjustments based on detected sleep events.

8 Device Alert

9 Troubleshooting

10 Cleaning and Maintaining

11 Maintenance

13 Specification

Resvent iBreeze 20A Specifications

General IconGeneral
Pressure Range4-20 cmH2O
HumidificationIntegrated heated humidifier
Data StorageSD card
DisplayLCD screen
Power Supply100-240 V AC, 50/60 Hz
Ramp Time0 to 45 minutes
Sound LevelLess than 30 dBA