How to fix Richmar Medical Equipment displays that fail to light up?
S
Sean LopezAug 4, 2025
If the displays on your Richmar Medical Equipment fail to light up, it may be due to adapter contact failure. Ensure the adapter is connected and check that all contacts are in place, not broken, and properly connected.
K
Kari SmithAug 11, 2025
Why is Richmar Medical Equipment stimulation uncomfortable?
M
Megan MccartyAug 11, 2025
Uncomfortable stimulation from Richmar Medical Equipment can occur for several reasons: the intensity might be too high (decrease intensity); the electrodes might be too close together (reposition the electrodes); the electrodes or lead wires might be damaged or worn (replace them); or the electrode active area size might be too small (replace electrodes with ones that have an active area no less than 2 inches).
S
Sharon GibsonAug 19, 2025
What to do if Richmar Medical Equipment stimulation is weak?
A
Amanda SolisAug 19, 2025
If the stimulation from your Richmar Medical Equipment is weak, there could be a few reasons. The electrodes may be dried out or contaminated (replace them), or they may not be placed correctly (electrodes must be a minimum of 1.5 inches apart and no more than 6 inches per channel). Also, the lead wires might be old, worn, or damaged (replace them).
A
Ann RodriguezAug 26, 2025
Why does Richmar Medical Equipment stimulation stop?
W
William GonzalezAug 27, 2025
Stimulation from Richmar Medical Equipment might stop due to poor electrode contact (reapply electrodes, secure firmly) or damaged/worn electrodes or lead wires (replace them).