RIGOL Appendix
4 DG4000 Performance Verification Guide
AC Flatness Test
Channel: CH1
Setting
Frequency
Settings
Measurement
Values
specifications Pass/Fail
[1]
1 kHz
Amplitude:
500 mVpp
Offset: 0 V
DC
Impedance:
50 Ω
P
ref
= -- --
10 MHz ±0.1 dB
60 MHz ±0.2 dB
100 MHz
[2]
±0.4 dB
160 MHz
[3]
±0.8 dB
Channel: CH2
Setting
Frequency
Settings
Measurement
Values
specifications Pass/Fail
[1]
1 kHz
Amplitude:
500 mVpp
Offset: 0 V
DC
Impedance:
50 Ω
P
ref
= -- --
10 MHz ±0.1 dB
60 MHz ±0.2 dB
100 MHz
[2]
±0.4 dB
160 MHz
[3]
±0.8 dB
Note
[1]
: Check if the result of “Measurement Value– P
ref
exceeds the corresponding specification.
Note
[2]
: Only for DG4202, DG4162 and DG4102.
Note
[3]
: Only for DG4202 and DG4162.