Appendix RIGOL
DG4000 Performance Verification Guide
5
Harmonic Distortion Test
Channel: CH1
Setting
Frequency
Settings Measurement Values specifications Pass/Fail
[1]
1 MHz
Waveform:
Sine
Amplitude:
0 dBm
Offset:
0 V
DC
1(fundamental wave):
<-60 dBc
--
2:
10 MHz
1(fundamental wave):
<-55 dBc
--
2:
100 MHz
[2]
1(fundamental wave):
<-50 dBc
--
2:
160 MHz
[3]
1(fundamental wave):
<-40 dBc
--
2:
Channel: CH2
Setting
Frequency
Settings Measurement Values specifications Pass/Fail
[1]
1 MHz
Waveform:
Sine
Amplitude:
0 dBm
Offset:
0 V
DC
1(fundamental wave):
<-60 dBc
--
2:
10 MHz
1(fundamental wave):
<-55 dBc
--
2:
100 MHz
[2]
1(fundamental wave):
<-50 dBc
--
2:
160 MHz
[3]
1(fundamental wave):
<-40 dBc
--
2:
Note
[1]
: Check if the result of “the measurement value of the second harmonic the measurement value of the
fundamental wave” exceeds the corresponding specification.
Note
[2]
: Only for DG4202, DG4162 and DG4102.
Note
[3]
: Only for DG4202 and DG4162.