Appendix Test Record Form RIGOL
DM3058/DM3058E Performance Verification Guide
1
Appendix Test Record Form
Zero Offset
Function
[1]
Range
Input
signal
Quick
Test
1-year Test Limit
Test Result
Upper
Limit
Lower
Limit
DC Voltage
200.000 mV
short
Q
[3]
8 μV
-8 μV
2.00000 V
short
60 μV
-60 μV
20.0000 V
short
800 μV
-800 μV
200.000 V
short
6 mV
-6 mV
1000.00 V
short
30 mV
-30 mV
DC Current
200.000 μA open Q 10 nA -10 nA
2.00000 mA open 100 nA -100 nA
20.0000 mA
open
Q
4 μA
-4 μA
200.000 mA
open
16 μA
-16 μA
2.00000 A open 400 μA -400 μA
10.0000 A
open
1 mA
-1 mA
Resistance
[2]
200.000 Ω
short
Q
10 mΩ
-10 mΩ
2.00000 kΩ
short
60 mΩ
-60 mΩ
20.0000 kΩ
short
600 mΩ
-600 mΩ
200.000 kΩ short 6 Ω -6 Ω
2.00000 MΩ
short
Q
80 Ω
-80Ω
10.0000 MΩ
short
300 Ω
-300 Ω
100.000 MΩ
short
4 kΩ
-4 kΩ
Note
[1]
: The reading rate is set to "Slow".
Note
[2]
: Specifications are for the 4-wire resistance function or the 2-wire resistance function using
“REL” operation. If the 2-wire resistance function is used without "REL" operation, add 0.2 Ω
additional errors.
Note
[3]
: Q marks the optional quick test points.