Appendix Test Record Form RIGOL
DM3058/DM3058E Performance Verification Guide
AC Voltage Gain
Range
[1]
Test
Signal
Input
Frequency
Quick
Test
1-year Test Limit
Test
Result
Upper Limit Lower Limit
200.000 mV
200.000 mV
200 mV 50 kHz 202.100 mV 197.900 mV
2.00000 V
2 V 20 kHz Q 2.00500 V 1.99500 V
2 V 50 kHz Q 2.02100 V 1.97900 V
2 V 100 kHz Q 2.06100 V 1.93900 V
20.0000 V
200.000 V
750.00 V
Note
[1]
: The reading rate is set to "Slow".
Note
[2]
: Q marks the optional quick test points.