RIGOL Appendix Test Record Form
4 DM3058/DM3058E Performance Verification Guid
AC Current Gain
Range
[1]
Test
Signal
Input
Frequency
Quick
Test
1-year Test Limit
Test
Result
Upper Limit Lower Limit
20.0000 mA
1 mA
1 kHz
Q
[3]
1.0450 mA
0.9550 mA
20 mA
20 Hz
20.3200 mA
19.6800 mA
20 mA
45 Hz
20.1200 mA
19.8800 mA
20 mA
2 kHz
Q
20.1200 mA
19.8800 mA
20 mA
10 kHz
20.5400 mA
19.4600 mA
200.000 mA
10 mA
1 kHz
10.430 mA
9.570 mA
200 mA
20 Hz
Q
203.200 mA
196.800 mA
200 mA
45 Hz
Q
200.800 mA
199.200 mA
200 mA 2 kHz Q 200.800 mA 199.200 mA
200 mA 10 kHz Q 205.400 mA 194.600 mA
2.00000 A
100 mA
1 kHz
0.10650 A
0.09350 A
2 A
20 Hz
2.03400 A
1.96600 A
2 A
45 Hz
2.01400 A
1.98600 A
2 A
2 kHz
Q
2.01400 A
1.98600 A
2 A
10 kHz
2.05400 A
1.94600 A
10.0000 A
[2]
500 mA
1 kHz
0.5275 A
0.4725 A
10 A
20 Hz
10.1650 A
9.8350 A
10 A
45 Hz
10.0650 A
9.9350 A
10 A
2 kHz
Q
10.0650 A
9.9350 A
10 A 5 kHz 10.2700 A 9.7300 A
Note
[1]
: The reading rate is set to "Slow".
Note
[2]
: For continuous currents greater than DC 7A or AC RMS 7A, close the circuit for 30 seconds
and then open the circuit for 30 seconds.
Note
[3]
: Q marks the optional quick test points.