RIGOL Chapter 2 Performance Verification Test
2-2 DM3068 Performance Verification Guide
2.00000mA open ±60nA ±60nA ±60nA
20.0000mA open Q ±2.4μA ±3μA ±3μA
200.000mA open ±4μA ±6μA ±6μA
2.00000A open Q ±400μA ±400μA ±400μA
10.0000A open ±1mA ±1mA ±1mA
Resistance
[2]
200.000Ω short ±6mΩ ±8mΩ ±8mΩ
2.00000kΩ short ±10mΩ ±20mΩ ±20mΩ
20.0000kΩ short Q ±100mΩ ±200mΩ ±200mΩ
200.000kΩ short ±1Ω ±2Ω ±2Ω
1.00000MΩ short Q ±10Ω ±10Ω ±10Ω
10.0000MΩ short ±100Ω ±100Ω ±100Ω
100.000MΩ short ±10kΩ ±10kΩ ±10kΩ
Note
[1]
: The integration time is set to 100PLC.
Note
[2]
: Specifications are for the 4-wire resistance function or the 2-wire resistance function using “REL” operation.
If the 2-wire resistance function is used without "REL" operation, add 0.2Ω additional errors.
Note
[3]
: Q marks the optional quick test points.
Gain Test
The gain test inspects the accuracy of the multimeter under full range. It is necessary only when a
regular gain offset occurs to a function and range.
1. The DC gain test procedures
(1) Make sure you have carefully read the "Test Conditions".
(2) Connect the input terminals of the multimeter with the calibrator.
(3) Test the functions under different ranges specified in Table 2-2 one by one. Set the
integration time to 100PLC. Unless otherwise noted, turn off all the math operations.
(4) Input the signals specified in Table 2-2 using the calibrator and compare the test results
with the limits in the table (make sure the output of the calibrator is built adequately).
Table 2-2 DC gain test items
Function
[1]
Range Test Signal
Quick
Test
[4]
Allowable Error Range
24 hours 90 days 1 year
DC Voltage 200.0000mV 200mV ±8μV ±11μV ±13μV
200.0000mV -200mV ±8μV ±11μV ±13μV
2.000000V 2V Q ±40μV ±52μV ±82μV
2.000000V -2V ±40μV ±52μV ±82μV
20.00000V 20V Q ±480μV ±700μV ±900μV
20.00000V -20V ±480μV ±700μV ±900μV
200.0000V 200V Q ±5.2mV ±9.2mV ±11.2mV
200.0000V -200V ±5.2mV ±9.2mV ±11.2mV
1000.000V 1000V Q ±26mV ±50mV ±65mV
1000.000V -1000V ±26mV ±50mV ±65mV