Appendix RIGOL
DM3068 Performance Verification Guide
Appendix Test Record Form
Zero Offset
Function
[1]
Range Input Signal
Quick
Test
[3]
Allowable Error Range Test Result
24 hours 90 days 1 year 24 hours 90 days 1 year
DC Voltage 200.0000mV short Q ±4μV ±5μV ±5μV
2.000000V short ±10μV ±12μV ±12μV
20.00000V short ±80μV ±100μV ±100μV
200.0000V short Q ±1.2mV ±1.2mV ±1.2mV
1000.000V short ±6mV ±10mV ±10mV
DC Current 200.0000μA open Q ±24nA ±30nA ±30nA
2.000000mA open ±60nA ±60nA ±60nA
20.00000mA open Q ±2.4μA ±3μA ±3μA
200.0000mA open ±4μA ±6μA ±6μA
2.000000A open Q ±400μA ±400μA ±400μA
10.00000A open ±1mA ±1mA ±1mA
Resistance
[2]
200.0000Ω short ±6mΩ ±8mΩ ±8mΩ
2.000000kΩ short ±10mΩ ±20mΩ ±20mΩ
20.00000kΩ short Q ±100mΩ ±200mΩ ±200mΩ
200.0000kΩ short ±1Ω ±2Ω ±2Ω
1.000000MΩ short Q ±10Ω ±10Ω ±10Ω
10.00000MΩ short ±100Ω ±100Ω ±100Ω
100.0000MΩ short ±10kΩ ±10kΩ ±10kΩ
Note
[1]
: The integration time is set to 100PLC.
Note
[2]
: Specifications are for the 4-wire resistance function or the 2-wire resistance function using “REL” operation. If the 2-wire resistance function is used without "REL"
operation, add 0.2 Ω additional errors.
Note
[3]
: Q marks the optional quick test points.