RIGOL Chapter 2 Constant Voltage Tests
DP1308A Performance Verification Manual
2-6
9. Compare the measurement results with the specification in “Appendix B:
Specifications”. The measurement results should be no greater than 0.01%.
10. Repeat steps 1 to 9 to test the CV linear adjustment rates of the other two output
channels.
Table 2-3 Setting and specification of CV linear adjustment rate test
Channel +6V/5A +25V/1A -25V/1A
Voltage (V) 6 25 -25
Current (A) 5.25 1.05 1.05
Specification <0.01%+2mV