RIGOL Appendix
DP1308A Performance Verification Manual
4
Appendix B: Specifications
The following parameters can be guaranteed only when the instrument has been
operating under the specified operation temperature for more than 30 minutes.
Note: unless otherwise noticed, all the specifications are applicable to the outputs of
the three channels.
Model DP1308A
Channel +6V +25V -25V
DC Output (0
to 40
)
Voltage 0 to +6V 0 to +25V 0 to -25V
Current 0 to 5A 0 to 1A 0 to 1A
Overvoltage Protection 0.1V to 6.5V 0.1V to 27V -0.1V to -27V
Overcurrent Protection 0.1A to 5.5A 0.1A to 1.2A 0.1A to 1.2A
Load Adjustment Rate ± (output percentage + offset)
Voltage <0.01%+2mV
Current <0.01%+250μA
Linear Adjustment Rate ± (output percentage + offset)
Voltage <0.01%+2mV
Current <0.01%+250μA
Ripple and Noise (20 Hz to 20 MHz)
Normal Mode Voltage <350μVrms/2mVpp
Normal Mode Current <2mArms <500μArms
Common Mode Current <1.5μArms
Year Accuracy
[1]
(25
±5
) ± (output percentage + offset)
Programming
Voltage 0.1%+5mV 0.05%+20mV
Current 0.2%+10mA 0.15%+4mA
Readback
Voltage 0.1%+5mV 0.05%+10mV
Current 0.2%+10mA 0.15%+4mA
Resolution
Programming 0.5mV/0.5mA 1.5mV/0.1mA
Readback 0.5mV/0.5mA 1.5mV/0.1mA
Electric Meter 1mV/1mA 10mV/1mA
Transient Response Time