EasyManuals Logo

Rigol DP831A Programming Guide

Rigol DP831A
134 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #32 background imageLoading...
Page #32 background image
RIGOL Chapter 2 Command System
2-14 DP800 Programming Guide
*SAV
Syntax *SAV {1|2|3|4|5|6|7|8|9|10}
Description Save the current system state to the specified storage location using the default
name. The default name is RIGOLn.RSF; n corresponds to the number of the
storage location.
Parameter
Name Type Range Default
{1|2|3|4|5|6|7|8|9|10} Discrete 1|2|3|4|5|6|7|8|9|10 None
Explanation The power supply provides 10 storage locations (numbered 1 to 10) for
instrument states.
If the specified storage location already contains a state file, this command will
directly overwrite the original file. If the state file stored in the specified
storage location is locked (refer to the
:MEMory[:STATe]:LOCK command), this
command will not overwrite the original file and the storage operation is
invalid.
Example *SAV 5 /*Save the current instrument state to storage location 5 with the
filename RIGOL5.RSF*/
Related
Command
:MEMory[:STATe]:STORe
*TST?
Syntax *TST?
Description Query the self-test results of the instrument.
Explanation The power supply executes self-test at start-up. This command queries the self-test
results (including TopBoard, BottomBoard and fan).
Return Format The query returns the self-test results of TopBoard, BottomBoard and fan
respectively, for example, TopBoard:PASS,BottomBoardPASS,Fan:PASS.
Related
Commands
:SYSTem:SELF:TEST:BOARD?
:SYSTem:SELF:TEST:FAN?

Table of Contents

Other manuals for Rigol DP831A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Rigol DP831A and is the answer not in the manual?

Rigol DP831A Specifications

General IconGeneral
BrandRigol
ModelDP831A
CategoryPower Supply
LanguageEnglish

Related product manuals