EasyManuals Logo

Rigol DS1000Z-E Series Performance Verification Guide

Rigol DS1000Z-E Series
26 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #23 background imageLoading...
Page #23 background image
Chapter 2 Performance Verification Test RIGOL
DS1000Z-E Performance Verification Guide
2-11
Zero Point Offset Test
Zero point offset is defined as the offset of the crossing point of the waveform and trigger level line
relative to the trigger position, as shown in the figure below.
Specification
Zero Point Offset
Specification
2.5 ns
Test Connection Diagram
Figure 2-6 Zero Point Offset Test Connection Diagram
Test Procedures
1. Connect the active signal terminal of Fluke 9500B to CH1 of the oscilloscope via a 50 Ω impedance
adapter, as shown in the figure above.
2. Turn on Fluke 9500B and set its impedance to 50 Ω.
3. Output a fast edge signal with 500 ps rise time and 1.2 V amplitude via Fluke 9500B.
4. Configure the oscilloscope:
1) Press CH1 in the vertical control area (VERTICAL) on the front panel to turn on CH1.
2) Press CH1 Probe to set the probe attenuation ratio to "1X".
3) Rotate VERTICAL SCALE to set the vertical scale to 200 mV/div.
4) Rotate HORIZONTAL SCALE to set the horizontal time base to 5 ns/div.
5) Rotate HORIZONTAL POSITION and VERTICAL POSITION respectively to
Fluke 9500B
DS1000Z-E
50 Ω Impedance Adapter

Other manuals for Rigol DS1000Z-E Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Rigol DS1000Z-E Series and is the answer not in the manual?

Rigol DS1000Z-E Series Specifications

General IconGeneral
BrandRigol
ModelDS1000Z-E Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals