EasyManuals Logo

Rigol DS1074B Service Guide

Rigol DS1074B
75 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #40 background imageLoading...
Page #40 background image
RIGOL Chapter 3 Performance Test
DS1000B Service Guide
3-10
Test Record Form (continue)
Channel
Vertical
Scale
Test Result
Limit Pass/Fail
Vavg1 Vavg2
Calculation
Result
[1]
CH4
2 mV/div
≤ 4%
5 mV/div
10 mV/div
≤ 3%
20 mV/div
50 mV/div
100 mV/div
200 mV/div
500 mV/div
1 V/div
2 V/div
5 V/div
10 V/div
Note
[1]
: The calculation formula is |(Vavg1 - Vavg2) - (Vout1 - Vout2)|/Full Scale × 100%; wherein,
Vout1 and Vout2 are 3 × the current vertical scale and -3 × the current vertical scale respectively.