RIGOL Chapter 1 Quick Start
1-8 MSO1000Z/DS1000Z User’s Guide
Probe Compensation
When the probes are used for the first time, you should compensate the probes to
make them match the input channels of the oscilloscope. Non-compensated or
poorly compensated probes may cause measurement inaccuracy or error. The probe
compensation procedures are as follows.
1. Perform steps 1, 2, 3 and 4 of "Function Inspection".
2. Check the displayed waveforms and compare them with the following figures.
Figure 1-9 Probe Compensation
3. Use a nonmetallic driver to adjust the low-frequency compensation adjustment
hole on the probe until the displayed waveform is as the "Perfectly
compensated" in the figure above.
Over compensated Perfectly compensated Under compensated
www.GlobalTestSupply.com
Find Quality Products Online at: sales@GlobalTestSupply.com

Table of Contents

Other manuals for Rigol DS1104Z Plus

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Rigol DS1104Z Plus and is the answer not in the manual?

Rigol DS1104Z Plus Specifications

General IconGeneral
Bandwidth100 MHz
Channels4
Sample Rate1 GSa/s
Memory Depth24 Mpts
Waveform Capture Rate30, 000 wfms/s
Vertical Sensitivity1 mV/div to 10 V/div
Timebase Range5 ns/div to 50 s/div
ConnectivityUSB Host, USB Device, LAN, AUX
Vertical Resolution8 bits
Power Supply100-240 V AC, 45-440 Hz
Display7 inch TFT LCD
Trigger TypesEdge, Pulse, Video, Slope
Input Impedance1 MΩ ± 2%

Related product manuals