RIGOL Chapter 2 Performance Verification Test
DSA1000A/DSA1000 Performance Verification Guide
2-4
Test Record Form
Center
Frequency
Maximum
Peak
Minimum
Peak
Average
[1]
Limit
[2]
Pass/Fail
Preamplifier off
103 kHz
100 kHz~10 MHz:
10 MHz~2.5 GHz:
2.5 GHz~3 GHz:
1.04 MHz
50.04 MHz
504 MHz
1.904 GHz
2.494 GHz
Preamplifier on
103 kHz
100 kHz~1 MHz:
1 MHz~10 MHz:
10 MHz~2.5 GHz:
2.5 GHz~3 GHz:
1.04 MHz
50.04 MHz
504 MHz
1.904 GHz
2.494 GHz
Note:
[1] Average = (Maximum Peak + Minimum Peak)/2 (dimensionless calculation)
[2] Fill in the corresponding specification according to your instrument.