RIGOL Chapter 2 Performance Verification Test
DSA1000A/DSA1000 Performance Verification Guide
2-14
4. Press Single and wait for the instrument to finish a sweep, press Peak to find the maximum
peak. Then, set the cursor type to Delta. Set the center frequency to 2 GHz, press Single and
wait for the instrument to finish a sweep, press Peak and record the delta result.
5. Second Harmonic Distortion=-20 dBm - Delta (-20 dBm is the level input into the mixer)
and compare the calculation result with the specification.
Test Record Form
Output Frequency of
Signal Generator
Delta Calculation Result
[1]
Limit Pass/Fail
1 GHz > 35 dBm
Note: [1] Calculation Result = -20 dBm - Delta (-20 dBm is the level input into the mixer).