RIGOL Chapter 1 Disassemble and Assemble
1-12 MSO1000Z/DS1000Z Service Guide
and take off the shield cover gently.
(4) Remove the screw () fixing the signal source board using the 5mm
hexagon socket, and take off the signal source board gently.
(5) Remove the screw () fixing the main board and front metal plate using
the 5mm hexagon socket.
(6) Remove the 9 screws () fixing the main board and front metal plate using
the screwdriver (T10).
(7) Take off the main board gently.
2. To Disassemble the Main Board of MSO1000Z/DS1000Z
Part Explanation:
10 screws (torx pan head screw assembly M3*6) fixing the main board and
front metal plate (please remove them according to their actual positions).
Disassemble Steps:
(1) Remove the 4 nuts and washers at the BNC ports on the front metal plate
using the BNC socket.
(2) Remove the 10 screws () fixing the main board and front metal plate
using the screwdriver (T10).
(3) Take off the main board gently.