- 45 -
AIR CAL
Press the ENTER switch.
The current concentration readings of the gases are
displayed. Press the AIR switch when they are
stabilized.
When the AIR switch is pressed, HOLD AIR is
displayed. Keep pressing the switch until RELEASE
is displayed.
Release the AIR switch.
* Only on TYPE-A and TYPE-E, a 30-second
countdown is started. No countdown is started on
other types.
After the fresh air adjustment is successfully
completed and END is displayed, the gas monitor
automatically returns to the menu.