- 41 -
7
Calibration Mode
Calibrate sensors using a calibration gas at least once every six months. A calibration consists of a zero
adjustment and a span adjustment.
Calibration requires a demand flow regulator, calibration tubing, and calibration gas.
To enter Calibration Mode, while the instrument is in normal operation, press and hold the
button then press and hold the DISPLAY button. Release both buttons when you hear a beep.
Calibration Mode’s menu items are shown below.
Perform the fresh air adjustment.
Zero adjustment for
high-concentration
combustible gases
Perform the zero adjustment for
high-concentration combustible
gas sensor.
Simultaneous span
adjustment for all channels
Perform the span adjustment
simultaneously for all channels.
Span adjustment for each
channel
Perform the span adjustment for
each channel.
Exit the calibration mode and
return to the detection mode.
7-1. Preparing for Calibration
Required equipment/material
Zero adjustment gas (100% N2) for high-concentration combustible gases
Span adjustment gas (NOTE: Calibration gas for the high concentration combustible gas sensor must be in a
balance of N2.)
Demand flow regulator
Calibration tubing
Stopwatch
Connection
Connect the equipment as shown below to perform a calibration.
Gas sampling probe
(with a tube)
CAUTION
Do not use a lighter gas to check the sensitivity of the gas monitor. A constituent of the lighter gas
may deteriorate the sensor performances.
Demand flow
regulator
screwed into gas
cylinder