What to do if RKI Instruments Gas Detectors show abnormal readings?
- DDavid BenjaminAug 13, 2025
If the LCD shows abnormally high or low readings on your RKI Instruments Gas Detectors, while other instruments do not, try these steps: First, replace the filters. If that doesn't solve the problem, recalibrate the unit. If the issue persists, replace the sensor for the affected channel(s) and calibrate those channels again. If you're still experiencing difficulties, further assistance from RKI Instruments, Inc. is recommended.