GX-2012
Operators Manual
Part Number: 71-0335
Revision: C
Released: 7/23/15

Table of Contents

Other manuals for RKI Instruments GX-2012

Questions and Answers

  • C
    Cindy JonesAug 18, 2025
    Why do Measuring Instruments by RKI Instruments show abnormally high or low readings?
    • J
      Jesse StevensAug 18, 2025
      Abnormally high or low readings on your RKI Instruments Measuring Instruments can occur if the LEL or CO filters need replacement. If the issue persists, recalibrate the unit or replace the sensor for the affected channel(s) and then calibrate the channel(s).
  • M
    Margaret JenningsAug 20, 2025
    What to do if auto calibration fails on RKI Instruments GX-2012?
    • D
      Diana BoothAug 20, 2025
      If auto calibration fails on your RKI Instruments Measuring Instruments, check the calibration tubing for leaks and bad connections. Also, ensure that the GX-2012 has been properly set up for calibration. Verify the calibration cylinder contains enough fresh test sample and change the CO filter if it is saturated. If the problem persists, replace the sensor(s).
  • T
    Todd GregoryAug 22, 2025
    What to do if RKI Instruments GX-2012 Measuring Instruments indicate flow failure?
    • E
      Emily HaneyAug 23, 2025
      If your RKI Instruments Measuring Instruments indicate a flow failure, try the following: * Change the cotton filter in the probe or filter holder. * Inspect the sample hose for kinks or obstructions and replace if necessary. * Inspect the hydrophobic filter disks and the wire mesh disk and replace if necessary. * Perform a low flow setpoint adjustment.
  • R
    Raven WestAug 26, 2025
    What to do if RKI Instruments GX-2012 display 'FAIL SYSTEM'?
    • D
      Daniel FowlerAug 26, 2025
      If your RKI Instruments Measuring Instruments display 'FAIL SYSTEM' during startup, it indicates a circuit fault. If the error code is 031, indicating a FLASH memory failure, press and release the RESET SILENCE button to continue to measuring mode keeping in mind that there will be no datalogging function. If the error code is anything else, the instrument cannot be used.
  • D
    donna77Aug 28, 2025
    How to fix 'FAIL CLOCK' error on RKI Instruments GX-2012?
    • C
      Courtney MackAug 28, 2025
      If your RKI Instruments Measuring Instruments display 'FAIL CLOCK' during startup, it means an internal clock failure has occurred. Press and release the RESET SILENCE button to use the unit temporarily. Attempt to set the date in the DATE menu item in Maintenance Mode.
  • M
    Michael RivasAug 30, 2025
    How to turn on RKI Instruments GX-2012 if the LCD is blank?
    • J
      jenniferwilsonAug 30, 2025
      If the LCD on your RKI Instruments Measuring Instruments is blank, first ensure the unit is turned on by pressing and briefly holding the POWER ENTER button. If it still doesn't turn on, the alkaline batteries may need replacement, or the lithium ion battery pack may need recharging. Replace the alkaline batteries or recharge the lithium ion battery pack.
  • B
    billydavisSep 1, 2025
    Why does my RKI Instruments Security Sensors show incorrect readings?
    • C
      colleen72Sep 1, 2025
      If your RKI Instruments Security Sensors unit shows abnormally high or low readings compared to other gas detection instruments, it could be due to several reasons. First, the LEL filter may need replacement if the abnormal reading is on the LEL channel. If the abnormal reading is on the CO channel, replace the charcoal filter. Also, recalibrate the unit, or replace the sensor for the affected channel(s) and calibrate the affected channel(s).
  • D
    Diana Craig MDSep 4, 2025
    What to do if RKI Instruments Security Sensors unit indicates flow failure?
    • M
      Michelle TranSep 4, 2025
      If your RKI Instruments Security Sensors unit indicates flow failure, and pressing the RESET SILENCE button doesn't resolve it, check the following: * The cotton filter in the probe or filter holder may be dirty and clogged, so change it. * The sample hose may have a kink or obstruction, so inspect and replace if necessary. * The hydrophobic filter disks or wire mesh disk may be dirty and clogged, so inspect and replace if necessary. * The pump may be malfunctioning, so perform a low flow setpoint adjustment.
  • A
    Andrea PenaSep 6, 2025
    Why does auto calibration fail on RKI Instruments Security Sensors?
    • J
      Julian MosleySep 7, 2025
      If auto calibration fails or you're unable to set the response readings during single calibration on your RKI Instruments Security Sensors, it could be due to several factors. Check all calibration tubing for leaks or bad connections. The CO filter might be saturated, so replace it. Ensure the GX-2012 has been properly set up for calibration and verify that the calibration cylinder contains an adequate supply of fresh test sample. If the issue persists, the sensor for the affected channel(s) may need replacement.
  • C
    claymarkSep 9, 2025
    How to fix a blank LCD on RKI Instruments Security Sensors?
    • C
      Carrie AdamsSep 9, 2025
      If the LCD is blank on your RKI Instruments Security Sensors, it could be due to a few reasons. The unit may be turned off, so press and briefly hold the POWER ENTER button to turn it on. Alternatively, the alkaline batteries may need replacement, or the lithium ion battery pack may need recharging. Replace the alkaline batteries or recharge the lithium ion battery pack.

Summary

Introduction to GX-2012 Gas Monitor

GX-2012 Overview and Features

Provides a general overview and key features of the GX-2012 gas monitor.

GX-2012 Applications and Specifications

Details the applications and technical specifications of the device.

GX-2012 Instrument Description

Physical Components of the GX-2012

Describes the case, display, buttons, and ports of the instrument.

Gas Sensor Technology and Functionality

Explains the different types of sensors and how they detect gases.

Standard and Optional Accessories Guide

Lists and describes accessories for the GX-2012.

Operating the GX-2012 in Normal Mode

Normal Mode Startup and Basic Operation

Guides users through powering on and initial setup in Normal Mode.

Gas Monitoring and Alarm Management

Explains how to monitor gases and respond to alarms.

Navigating Display Mode Screens

Details various display screens and their functions.

GX-2012 Bar Hole Mode Operations

Bar Hole Mode Setup and Testing

Explains how to use the instrument for bar hole testing.

GX-2012 Leak Check Mode Procedures

Leak Check Mode Startup and Testing

Guides through leak testing and related features.

GX-2012 Maintenance and Calibration

Maintenance Mode Settings and Navigation

Explains how to access and adjust instrument settings.

Calibration and Bump Test Procedures

Covers essential calibration and bump test operations.

GX-2012 Maintenance and Troubleshooting

Troubleshooting Common Issues and Solutions

Provides guidance for diagnosing and resolving problems.

Battery Replacement and Recharging

Details procedures for maintaining the instrument's power source.

Replacing Filters and Sensors

Explains how to replace consumable parts.

Appendix A: Sample Bag Calibration and Bump Testing

Using the Sample Bag Calibration Kit

Describes how to use a sample bag for calibration.

RKI Instruments GX-2012 Specifications