52 Display Mode GX-2012 Operator’s Manual
2. Use the AIRS and (SHIFT)T buttons to scroll through the 128 possible
user IDs.
3. Press the POWER ENTER button to select a user ID.
Station ID Screen
This screen always appears in Display Mode, regardless of the ID DISP
setting in the Maintenance Mode menu. Each station ID can contain up to 16
characters but only the first 8 of those characters are visible in the instrument.
Be sure that all station IDs are distinguishable from each other by viewing
only the first 8 characters. For more information, see the GX-2012 Data
Logger Management Program manual. The bottom line of the screen
alternates between YES/ENT., the current station ID, and NO /DISP.
Uppercase letters, numbers, asterisks (*), dashes (-), and a blank space are
available characters. The GX-2012 can store 128 station IDs.
Use this screen to identify or change the station. The station ID provides a
way to identify the station during a data logging session. The station ID is
saved to the data logger for that session when you turn off the GX-2012 or
change stations during operation. This allows you to change the station ID
during operation and have each station ID that was used during an operating
session saved for the corresponding data.
NOTE: As shipped from the factory, the station ID on this screen is shown
as -- -- -- -- -- -- until you enter this screen and select a station ID.