138 Using Maintenance Mode GX-2012 Operator’s Manual
Updating the Bar Hole Sampling Time
The Bar Hole Sampling Time is the amount of time that the GX-2012
samples a bar hole when being used in Bar Hole Mode. The 3 selections are
30, 45, and 60 seconds. The factory setting is 30 seconds.
1. Use the AIRS or (SHIFT)T buttons to display the B.H. TIME menu
item.
2. Press and release the POWER ENTER button. The screen will indicate
the current setting.
3. Use the AIRS or (SHIFT)T button to display the desired setting.
4. Press and release the POWER ENTER button to save the setting and
return to the main menu.
Turning Portuguese On or Off
With PORTUGAL set to On, the GX-2012’s displayed language will
be Portuguese.
With PORTUGAL set to OFF (factory setting), the GX-2012’s
displayed language will be English.
MAINTENANCE
B.H. TIME
30
MAINTENANCE
B.H. TIME