EasyManua.ls Logo

RKI Instruments GX-3R - Entering Gas Select Mode; Tips for Using Gas Select Mode

RKI Instruments GX-3R
122 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
104 • Appendix B: Gas Select Mode GX-3R Operators Manual
Entering Gas Select Mode
WARNING: The GX-3R is not in operation as a gas detector while in Gas Select Mode.
1 . Take the GX-3R to a non-hazardous location and turn it off if it is on.
2 . Press and hold AIR, then press and hold POWER MODE. You will hear a beep after one
second. Continue to hold the buttons down.
3 . You will hear a second beep. Continue to hold the buttons down.
4 . When you hear a third beep, release the buttons.
5 . A password screen appears. The first digit flashes. The password is 2014.
6 . Use AIR to select each password number then press POWER MODE to save it and move
on to the next number. To go back a number, press and hold AIR and POWER MODE
for a few seconds. To reverse the direction of change (ie. from increasing to decreasing
or vice versa):
a. Press and hold AIR.
b. Immediately press POWER MODE and then release both buttons.
7 . The GAS COMB menu item displays.
8 . at the top of the screen indicates that the GX-3R is in Gas Select Mode.
9 . Use AIR to move through the Gas Select Mode menu items.
Tips for Using Gas Select Mode
To scroll from one menu item to the next, press and release AIR. To reverse the scrolling
direction:
a. Press and hold AIR.
b. Immediately press POWER/MODE and then release both buttons.
c. The scrolling direction returns to the original direction when you exit and reenter a
menu.
To skip an item when a question is asked, press and release AIR.
To enter an item and to save any changes, press and release POWER MODE.
PASSWORD
M 7:49
0 0 0 0
GAS COMB