12 • Chapter 2: Description GX-3R Operators Manual
Control Buttons
Two control buttons, AIR and POWER MODE, are located below the LCD.
Alarm LEDs
The six alarm LEDs around the edge of the case alert you to gas, low battery, and failure
alarms.
Buzzer
One solid-state electronic buzzer is located inside the case. Sound exits the case through a
hole in the middle front of the case. The buzzer sounds for gas alarms, malfunctions, low
battery voltage, and as an indicator during use of the GX-3R’s many display and adjustment
options.
Vibrator
A vibrating motor inside the GX-3R case vibrates for gas alarms, unit malfunctions, and as an
indicator during normal use of the various modes of the GX-3R.
NOTE: If STEALTH is set to ON, the vibrator only functions when VIB in the STEALTH
Gas Select Mode item is set to ON. See “Stealth and Vibrator Settings
(STEALTH)” on page 109.
Sensors
The GX-3R uses three sensors to monitor combustible gas, oxygen (O
2
), carbon monoxide
(CO), and hydrogen sulfide (H
2
S) simultaneously. The sensors are located inside the GX-3R
and are held in their sockets by the sensor retainer and bottom cover. The sensors use different
detection principles, as described below.
Combustible Gas Sensor
The combustible gas sensor detects combustible gas in the % LEL range using a catalytic
element. The element’s resistance changes based on the reaction of gas with oxygen. The
change in resistance affects the current flowing through the element. The GX-3R’s circuitry
amplifies the current, converts the current to a gas concentration, and displays the
concentration on the LCD.
Table 3: GX-3R Control Button Functions
Button Function(s)
AIR turns on LCD backlight
resets alarm condition if LATCHING is set to ON in Maintenance Mode
enters User Mode, Maintenance Mode, and Gas Select Mode when used
with POWER MODE button
activates the demand zero function (adjusts the GX-3R’s fresh air reading)
changes the value of a parameter available for adjustment
scrolls through parameter options
POWER MODE turns the GX-3R on and off
turns on LCD backlight
enters and scrolls through Display Mode
enters instructions into the GX-3R’s microprocessor
resets alarm condition if LATCHING is set to ON in Maintenance Mode
enters User Mode, Maintenance Mode, and Gas Select Mode when used
with AIR button

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the RKI Instruments GX-3R and is the answer not in the manual?

RKI Instruments GX-3R Specifications

General IconGeneral
BrandRKI Instruments
ModelGX-3R
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals