Setting the Confirmation Beep and
Non-Compliance Indicator (BEEP)
The BEEP menu item has 3 items: BEEP SEL, BEEP INT, and ESCAPE.
1 . While in User Mode, press AIR to scroll to BEEP.
2 . Press and release POWER MODE. The BEEP SEL menu item appears.
BEEP SEL
BEEP SEL defines what kind of confirmation or non-compliance indication you want to
occur in Measuring Mode. The available choices are:
OFF
(factory setting): The GX-3R does not provide a confirmation alert or non-compliance
indicator.
LED: The GX-3R’s LEDs double flash as often as defined by the BEEP INT parameter to
verify that the instrument is operating.
BUZZER: The GX-3R’s buzzer double beeps as often as defined by the BEEP INT
parameter to verify that the instrument is operating.
LED+BUZZ: The GX-3R’s LEDs double flash and the buzzer double beeps as often as
defined by the BEEP INT parameter to verify that the instrument is operating.
BUMP/CAL: If a bump test or a calibration is due and if BUMP.EXPD or CAL EXPD is set
to CONFIRM or NONE, the GX-3R’s LEDs double flash as often as defined by the BEEP
INT parameter to indicate a non-compliance. Once a bump test or calibration (depending on
which is due) is done, the LEDs stop flashing.
ALM ALRT
: If the instrument goes into any gas alarm, the LEDs double flash as often as
defined by the BEEP INT parameter to indicate a non-compliance. Once a successful bump
test or calibration is done, the LEDs stop flashing.
NOTE: Depending on the Need to get Bump Log setting (only accessed via SDM-3R
program), a data download may also be needed to clear the non-compliance.
B/C/ALM: The LEDs double flash to indicate a non-compliance if any of the following
happens.
a. BUMP.EXPD is set to CONFIRM or NONE and a bump test is due (cleared by
successful bump test).
b. CAL EXPD is set to CONFIRM or NONE and a calibration is due (cleared by
successful calibration).
BEEP
M 7:49
BEEP SEL
M 7:49

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the RKI Instruments GX-3R and is the answer not in the manual?

RKI Instruments GX-3R Specifications

General IconGeneral
BrandRKI Instruments
ModelGX-3R
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals