- 23 -
4-6. Modes
Details on each mode are provided as follows. (* Operations are slightly different depending on the type.)
Mode
Item
LCD display
Details
Detection
Mode
Concentration display
Normal state
Air Calibration
Mode
AIR CAL
Perform the zero adjustment.
Display and
Setting Mode
Combustible
Gas
Measurement
Range Setting
HC RANGE
Used to select a combustible gas
measurement range manually.
(TC sensor versions only)
Peak Display
PEAK
Display the maximum concentration (or
minimum concentration for oxygen)
detected during measurement from
power-on to the present.
STEL Value
Display
STEL
Display the STEL value after power-on.
(for versions with H2S and/or CO sensors)
TWA Value
Display
TWA
Display the TWA value after power-on.
(for versions with H2S and/or CO sensors)
Alarm Setpoint
Display
Alarm Test
ALARM-P
Display the full scale and alarm setpoint
values and check the alarm operations for
the settings displayed.
Pump
ON/OFF
Setting
PUMP OFF
Used to turn on/off the pump operations.
ID Setting
ID SELECT
Display an ID if it has been set in advance.
Also used to change or set an ID.
Log Data
Display
REC.DATA
Display data recorded to the manual
memory.