26 Warning messages
Biohazardous materials
Operator infection and injury Contact with system mechanisms, for example, sample
probes, reagent probes, rinse stations, the reaction cell
rinse unit, or with the system chassis or covers, may result
in personal injury and infection.
r Keep all covers and doors closed and in place while
the system is operating. An interlock system prevents
operation of moving parts when the covers are open.
r Always ensure that the system is powered off, in sleep
mode, or in standby mode, before you work with an
opened cover, for example, for cleaning or
maintenance.
r Do not open the covers while the system is performing
maintenance.
r Make sure the covers are fully open to avoid hitting
your head.
r Do not touch any parts of the analyzer other than
those specified.
r Never reach into the analyzer while parts are moving.
r Carefully observe all instructions given in this safety
manual.
Infectious samples Contact with samples containing material of human origin
may result in infection. All materials and mechanical
components associated with samples containing material
of human origin are potentially biohazardous.
r Follow good laboratory practices, especially when
working with biohazardous material.
r Keep all covers closed while the system is operating.
r Wear appropriate personal protective equipment.
r If any biohazardous material is spilled, wipe it up
immediately and apply laboratory disinfectant.
r If sample or waste comes into contact with your skin,
wash the affected area immediately with soap and
water and apply disinfectant.
Consult a physician.