Maintenance and Trouble Shooting
91
C
C
3572 Large Tip
missing
No large Tip was detected in
referenced channel number.
Check if tip tray has been
inserted and contains large
tip.
No large Tip was detected in
referenced channel number.
Check if tip tray has been
inserted and contains large
tip.
3573 Too much
sample
Larger volume than selected
was detected in the sample
tube in referenced channel
number.
Check and correct sample
volume in respective
channel, if needed.
Larger volume than selected
was detected in the sample
tube in referenced channel
number.
Check and correct sample
volume in respective
channel, if needed, and
restart run.
3580 Run aborted. Run was aborted by
operator.
Run was stopped during
purification, liquid waste
discard or leakage test due
to operator interaction,
obstacles in the movement
path or incorrect heating
temperature.
Repeat run. If problem
persists, contact Technical
Service.
3581 Run aborted
by error.
Run was aborted by an error. Run was aborted by an
error.
Repeat run. If problem
persists, contact Technical
Service.
3582 Obstacles on
stage.
Run was aborted due to
obstacles on the stage.
Run was aborted due to
obstacles on the stage.
Repeat run. If problem per-
sists, contact Technical
Service.
3583 Temperature
timeout error.
Process aborted due to time-
out error of heating unit
temperatures.
Process aborted due to
timeout error of heating unit
temperatures.
Restart system. If problem
persists, contact Technical
Service.
3584 New database
was created
frequently.
New database was created
frequently. Original database
was not found or not
functioning properly.
New database was created.
Original database was not
found or not functioning
properly.
If problem persists, contact
Technical Service.
3585 Unspecific
database
error.
Unspecific database error. Unspecific database error. Restart system. If problem
persists, contact Technical
Service.
3587 ID entry
missing in
Ch. X
Cartridge ID input is
missing.
Check correct position of the
cursor. Insert barcode ID
and proceed sample
ordering
ID input for cartridge is
missing.
Check correct position of
the cursor. Insert barcode
ID and proceed sample
ordering.
3588 IC ID entry
missing in Ch.
X
Cartridge ID input is
missing.
Check correct position of the
cursor. Insert barcode ID
and proceed sample orde-
ring
ID input for Internal Control
is missing.
Check correct position of
the cursor. Insert barcode
ID and proceed sample
ordering.
Error
Code
ID
Error
Message
Error Information
(displayed on
information screen)
Explanation Action
Error Codes and Trouble Shooting Guide
Instrument Error Codes
MagNA Pure Compact Manual.book Seite 91 Dienstag, 28. Februar 2006 4:16 16