EasyManua.ls Logo

Roche MagNA Pure Compact Instrument - Page 118

Roche MagNA Pure Compact Instrument
130 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
MagNA Pure Compact Operator’s Manual - Version 1.3
92
C
C
3589 Sample ID
missing in
Ch. X
Cartridge ID input is
missing.
Check correct position of the
cursor. Insert barcode ID
and proceed sample
ordering
ID input for sample tube is
missing.
Check correct position of
the cursor. Insert barcode
ID and proceed sample
ordering.
3590 Elution ID
missing in
Ch. X
Cartridge ID input is mis-
sing.
Check correct position of the
cursor. Insert barcode ID
and proceed sample
ordering
ID input for elution tube is
missing.
Check correct position of
the cursor. Insert barcode
ID and proceed sample
ordering.
3607 Small Tip
missing
No small Tip was detected in
referenced channel number.
Check if tip tray has been
inserted and contains small
tip.
No small Tip was detected
in referenced channel num-
ber.
Check if tip tray has been
inserted and contains small
tip.
3608 Piercing Tool
missing
No Piercing Tool was
detected in referenced
channel number.
Check if tip tray has been
inserted and contains a
Piercing Tool.
No Piercing Tool was
detected in referenced
channel number.
Check if tip tray has been
inserted and contains a
Piercing Tool.
Error
Code
ID
Error
Message
Error Information
(displayed on
information screen)
Explanation Action
Error Codes and Trouble Shooting Guide
Instrument Error Codes
MagNA Pure Compact Manual.book Seite 92 Dienstag, 28. Februar 2006 4:16 16

Table of Contents

Related product manuals