R&S FSH  Putting into Operation 
Connectors on the R&S FSH 
Quick Start Guide 1309.6269.12 - 09  22 
1.4.2  Tracking Generator Output 
The tracking generator output is located on the top of the R&S FSH. 
 
The tracking generator output power of the R&S FSH is 0 dBm nominal. 
With an integrated step attenuator, you can reduce the tracking generator output power 
by a maximum of 40 dB in 1 dB steps. 
Risk of instrument damage 
To avoid damage to the tracking generator output, the reverse voltage up to the voltage 
stated on the R&S FSH housing must not be exceeded. 
 
1.4.3  Power Sensor Port 
The connector for power sensors is located on the top of the R&S FSH. 
 
The connector is configured especially for power sensors. It is used for both the power 
supply and data transfer via the power sensor interface. 
Additionally, you can connect accessories like the R&S TS-EMF isotropic antenna 
(order no. 1158.9295.13) to the power sensor port.