EasyManua.ls Logo

Rohde & Schwarz FSH4 - Measuring the Power with a Power Sensor

Rohde & Schwarz FSH4
77 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
R&S FSH Getting Started
Using a Power Sensor
Quick Start Guide 1309.6269.12 - 09 65
2.3.1 Measuring the Power with a Power Sensor
The power sensors R&S FSH-Z1 and R&S FSH-Z18 measure the power in the range
from 10 MHz to 8 GHz and 18 GHz respectively.
For more information about the characteristics of the R&S NRP power sensors, refer to
their datasheets.
Risk of damaging the power sensor
Because of high input power,
the continuous power applied to the power sensor’s input must not exceed
400 mW (26 dBm)
use an attenuator for measurements on highpower transmitters
However, brief power peaks 10 µs up to 1 W (30 dBm) are permissible.
Test setup
Connect the cable of the power sensor to the power sensor jack of the R&S FSH.
1 Power sensor R&S FSH-Z1 or R&S FSH-Z18
2 Power sensor connector (DUT)
3 Power sensor jack
Measuring the power
Press the MODE key.
Press the "Power Meter" softkey.
The R&S FSH switches its operating mode.
If the R&S FSH recognizes a power sensor, it sets up a connection via the interface
and after a few seconds shows the measured power.
If no power sensor has been connected or is not connected appropriately, the
R&S FSH shows nothing.

Table of Contents

Other manuals for Rohde & Schwarz FSH4

Related product manuals