R&S FSH Getting Started
Using the Network Analyzer
Quick Start Guide 1309.6269.12 - 09 63
2.2.3 Measuring Scalar Transmission
Scalar measurements are available for all R&S FSH models with a tracking generator.
Connect the DUT between the RF input and the tracking generator output.
Note that you need an R&S FSH with VSWR bridge to be able to measure the
transmission in forward and reverse direction. Models without VSWR bridge can
measure the transmission in reverse direction only.
Press the MEAS key.
Press the "Meas Mode" softkey.
Select the "Scalar" menu item.
The R&S FSH starts scalar measurements.
Press the FREQ key.
Enter the frequency of the DUT.
Press the SPAN key.
Enter an appropriate span.
As you have changed the frequency settings, you should calibrate the R&S FSH for
more accurate results.
Calibrate (normalize) the R&S FSH for scalar transmission measurements in the
direction you want (see above).
Reconnect the DUT.
The R&S FSH shows the magnitude of the transmission.
To evaluate the results, you can use, for example, markers.