R&S FSH Spectrum Analyzer Mode
Configuring Spectrum Measurements
Operating Manual 1173.6275.12 - 30 111
3.2.5 Working with Traces
The trace menu contains all functions available to customize the trace display.
3.2.5.1 Selecting the Trace Mode
The R&S FSH provides several trace modes. The trace mode defines the way the
R&S FSH writes the trace.
Clear/Write
In its default state, the R&S FSH overwrites the trace after each sweep.
You can apply all detectors in this mode.
Average
The trace is the result of the moving average over several sweeps.
The R&S FSH calculates the (moving) average of the power levels for each pixel
over a particular number of sweeps in the range from 2 to 999.
Averaging reduces the effects of noise, but has no effects on sine signals. Using
the trace averaging therefore is a good way to detect signals in the vicinity of
noise.
You can apply all detectors in this mode.
Max Hold
The trace shows the maximum power levels that have been measured at each
pixel.
To overwrite a max hold trace, change a parameter in a way that the results can
not be compared any more, e.g. the span. Using the max hold trace mode is a
good way to detect intermittent signals or the maximum values of fluctuating
signals, for example.
Using the max hold trace mode automatically activates the max peak detector.
Min Hold
The trace shows the minimum power levels that have been measured at each
pixel.
To overwrite a min hold trace, change a parameter in a way that the results can not
be compared any more, e.g. the span. Using the min hold trace mode is a good
way to highlight signals within noise or suppress intermittent signals.
Using the min hold trace mode automatically activates the min peak detector.
View
The view trace mode freezes the current trace and aborts the measurement.
Using the view trace mode is a good way to evaluate the trace, for example with
markers.