R&S FSH Network Analyzer Mode
Performing Scalar Measurements
Operating Manual 1173.6275.12 - 30 189
6.3.2 Measuring the Reflection
The example included here is based on a reflection measurement of the same two-port
filter used for the transmission measurement. The filter has a center frequency of
2060 MHz and a bandwidth of about 11 MHz.
You can skip the preset, scalar measurement selection and frequency settings if you
have already set up the R&S FSH for the measurement. You can also skip the
calibration, if you have performed the appropriate calibration already.
Preset the R&S FSH
Before starting the measurement procedure, preset the R&S FSH to restore the default
configuration and connect the filter between the measurement ports.
Press the PRESET key.
Connect the DUT.
Start scalar measurements
Press the MEAS key.
Press the "Meas Mode" softkey.
Select the "Scalar" menu item.
Select the type of reflection measurement
Press the "Result Display" softkey.
Select the "Reflection Port 1" menu item for reflection measurements on port 1 or
the "Reflection Port 2" menu item for reflection measurements on port 2.
Select the "Reflection Port 1 and Port 2" menu item for a simultaneous display of
both the reflection on port 1 and on port 2.
In case of a simultaneous display, the R&S FSH shows two traces with a different
color in the result display. One for the measurement on port 1 and one for the
measurement on port 2. A label in the result display shows the port a trace belongs
to.
Note that the simultaneous display of both ports is only possible with a high
accuracy calibration.