R&S FSH Distance-to-Fault Mode (R&S FSH-K41)
Configuring Cable and Antenna Tests
Operating Manual 1173.6275.12 - 30 228
7.2.4 Configuring the Tracking Generator
If your R&S FSH features a tracking generator, you can define several characteristics
of the generated signal.
Press the AMPT key.
Press the "TG Settings" softkey.
You can define the level of the generated signal as an absolute level or an
attenuation (value relative to 0 dBm).
Select the “TG Power” menu item to define an absolute level.
Select the “TG Attenuation” menu item to define a relative level.
An input of 10 dB, for example, corresponds to an attenuation of -10 dBm.
In addition, you can define an arithmetic level offset, for example to take external
attenuators into account.
Press the "TG Settings" softkey.
Select the “TG Power Offset” menu item to define a level offset.
Note the the level offset has no effect on the actual signal level.
Tracking generator output level
Using an output level of 0 dBm (or a receiver attenuation of 0 dB) for S12 or S21
measurements can cause an overload at the RF input. It is therefore recommended to
use an output level of -10 dBm (default) or lower for these measurements.
This especially applies to measurements on amplifiers.