R&S FSH Distance-to-Fault Mode (R&S FSH-K41)
Configuring Cable and Antenna Tests
Operating Manual 1173.6275.12 - 30 227
Adjusting the vertical axis automatically
The R&S FSH provides an automatic scaling routine that scales the vertical axis in a
way that the results fit ideally on the display. The R&S FSH does this by determining
the minimum and maximum trace values and scaling the vertical axis according to
these values.
Press the AMPT key.
Press the "Range" softkey.
Select the "Auto Scale" menu item.
The R&S FSH performs automatic adjustment of the vertical axis.
Setting the reference position
The reference position defines the position of the reference line in the diagram. The
reference position is a linear value between 0 and 10. Each value represents one
horizontal grid line of the diagram. 0 corresponds to the top grid line and 10
corresponds to the bottom grid line.
When you change the reference position, the R&S FSH also shifts the position of the
trace by the magnitude of the reference position change. It has no effect on the
reference value itself.
Press the AMPT key.
Press the "Ref Pos" softkey.
Enter the reference position you want.
The R&S FSH moves the trace accordingly.
7.2.3.2 Setting the Attenuation
The R&S FSH provides functions to attenuate the signal at the RF input.
Press the "RF Att/Amp/Imp" softkey.
Select the "Manual:" menu item or one of the automated attenuation modes.
Enter the attenuation of the signal at the RF input.
You can set an RF attenuation in the range from 0 dB to 40 dB in 5 dB steps.