R&S FSH Network Analyzer Mode
Evaluating the Results
Operating Manual 1173.6275.12 - 30 206
Changing the reference position
Changing the reference position moves the position of the reference value up or down.
The reference position also moves the trace up or down. The levels and the reference
value, however, remain the same. The default reference position is the grid line at the
top of the diagram. The reference position is indicated by a triangle at the vertical axis
( ).
The reference position is a number between 0 and 10 with 0 being the grid line at the
bottom and 10 being the grid line at the top of the diagram.
Press the AMPT key.
Press the "Ref Pos" softkey.
The R&S FSH opens an input field.
Enter the new position of the reference value.
The R&S FSH moves the trace accordingly and moves the triangle to the new
position.