R&S FSH Network Analyzer Mode
Vector Voltmeter (R&S FSH-K45)
Operating Manual 1173.6275.12 - 30 212
6.6.2.2 Performing Reflection Measurements
Calibrate the measurement for reflection measurements.
Press the "Calibration" softkey.
Select the "Reflection Port 1" menu item.
Perform calibration. For more information see "Performing Calibration".
Reconnect the DUT.
The R&S FSH shows the characteristics of the DUT (magnitude and phase) as
numerical values.
You can change the measurement configuration (e.g. the sweep time) or format
(see "Selecting the Measurement Format") without affecting the accuracy of the
measurement.
Note that for calibration to remain valid, frequency parameters, bandwidth and
attenuation have to remain the same.