R&S FSH Distance-to-Fault Mode (R&S FSH-K41)
Vector Voltmeter (R&S FSH-K45)
Operating Manual 1173.6275.12 - 30 215
For measurement on DUTs that need an external voltage supply (e.g. power
amplifiers), you can connect the supply voltage from a suitable AC power supply to
the BIAS Port 1 or use the internal bias.
By default, the R&S FSH is in spectrum analyzer mode after you have turned it on. To
switch to the DTF mode from another operating mode, proceed as follows.
Press the MODE key.
Press the "Distance-to-Fault" softkey.
The R&S FSH starts the DTF mode.
By default, the R&S FSH performs measurements on port 1. It however, also supports
measurements on port 2, but you have to configure those first.
Press the "Meas Mode" softkey.
Select the "DTF / Refl Measurement Port" menu item.
The R&S FSH opens a submenu.
Select either "Port 1" or "Port 2".
Screen layout of the cable and antenna analyzer
1 Cable model
2 Hardware settings
3 Marker information
4 Status line:
- S-matrix
- Calibration status
- Measurement format
5 Trace window
6 Marker (blue line)
7 Cable length information
8 Cable frequency information
9 Cable test softkey menu