R&S FSH Receiver Mode (R&S FSH-K43)
Selecting the Measurement Mode
Operating Manual 1173.6275.12 - 30 233
8.1.1.1 Defining the Receive Frequency
In single frequency measurement mode, the R&S FSH determines the power level of a
single frequency only. The available frequency range depends on the R&S FSH model
you are using.
Press the FREQ key.
Press the "Freq" softkey.
Enter the frequency you want to measure.
For a quick change of frequencies with the rotary knob, define a frequency stepsize.
Press the "Freq Stepsize" softkey.
Enter a frequency stepsize.
For measurements on systems that use channels instead of single frequencies, you
can also load a channel table and measure the channel power.
Press the "Channel Mode" softkey.
The R&S FSH opens a dialog box to select a channel table.
For measurements based upon a channel table, you can select a single channel
instead of a single receive frequency.
Press the "Channel" softkey.
Enter a channel number.
8.1.1.2 Customizing the Bargraph Aspects
The bargraph is a graphical representation of the power level with one dimension, the
power level of the receive freuqency. The R&S FSH provides several ways to
customize the aspects of the bargraph.
Selecting the Unit
By default, the unit the R&S FSH uses in receiver mode for the measured power levels
in general is dBµV. In addition to dBµV, the receiver mode provides other units as well.
Press the AMPT key.
Press the "Unit" softkey.
Select the unit you need.