Trying Out the Instrument
R&S
®
SMA100B
48Getting Started 1419.8620.02 ─ 08
To save the instrument settings in a file
We assume, a test configuration as described in Chapter 6.2, "Generating an RF
Frequency Sweep Signal", on page 45.
1. Press the [Setup] key on the front panel.
2. In the "Setup" menu, select "Settings > Save/Recall".
3. In the "Save/Recall" dialog, select "Operation Mode > Save".
4. Tap the "Filename", use the on-screen keyboard, and enter MyTestSignal.
5. Tap the "Save" button.
The file MyTestSignal.savrcltxt is stored in the default direc-
tory /var/user/.
To load saved instrument settings
You can restore the settings to the instrument at any time using the settings file.
1. Press the [Preset] button to restore the default instrument settings so you can
check that the stored user settings are restored afterwards.
Saving and Recalling Settings