Waveform analysis
R&S
®
RTM3000
136User Manual 1335.9090.02 ─ 09
7.5.4 Width search
The width search finds pulses with an exact pulse width, or pulses shorter or longer
than a given time, or pulses inside or outside the allowable time range. It is similar to
the width trigger.
[Search] > "Search Type" = "Width" > "Setup"
Polarity
Indicates the polarity of the pulse to be searched for.
Remote command:
SEARch:TRIGger:WIDTh:POLarity on page 490
Level
Sets the voltage level on which the pulse width is measured. To let the instrument set
the level, tap "Find Threshold".
Remote command:
SEARch:TRIGger:WIDTh:LEVel on page 490
Hysteresis
Sets a hysteresis range to the search level to avoid unwanted search results caused
by noise oscillation around the level. To let the instrument set the hysteresis, tap "Find
Threshold".
Remote command:
SEARch:TRIGger:WIDTh:LEVel:DELTa on page 490
Comparison
Sets the condition how the measured pulse width is compared with the given limits.
Search