Remote control commands
R&S
®
RTM3000
779User Manual 1335.9090.02 ─ 09
<n> *
Selects the frame.
Return values:
<StopTime> Default unit: s
Usage: Query only
17.14 Signal generation (option R&S RTM-B6)
17.14.1 Function generator
17.14.1.1 Basic settings of the function generator
WGENerator:FUNCtion...................................................................................................779
WGENerator:VOLTage....................................................................................................779
WGENerator:VOLTage:OFFSet....................................................................................... 780
WGENerator:FREQuency............................................................................................... 780
WGENerator:FUNCtion:PULSe:DCYCle...........................................................................780
WGENerator:TRIangle:SYMMetry....................................................................................780
WGENerator:FUNCtion:PULSe:ETIMe.............................................................................780
WGENerator:FUNCtion:EXPonential:POLarity.................................................................. 781
WGENerator:FUNCtion:RAMP:POLarity...........................................................................781
WGENerator:NOISe:ABSolute.........................................................................................781
WGENerator:NOISe:RELative.........................................................................................781
WGENerator:OUTPut:LOAD........................................................................................... 781
WGENerator:OUTPut[:ENABle]....................................................................................... 781
WGENerator:FUNCtion <Function>
Selects the function to be generated.
Parameters:
<Function> DC | SINusoid | SQUare | PULSe | TRIangle | RAMP | SINC |
ARBitrary | EXPonential
WGENerator:VOLTage <Amplitude>
Defines the amplitude value (peak-to-peak value) of the selected generator function.
Parameters:
<Amplitude> Numeric value
Range: 6.0000E-02 to 6.00000E+00
*RST: 5.0000E-01
Default unit: Vpp
Signal generation (option R&S RTM-B6)