Serial bus analysis
R&S
®
RTM3000
281User Manual 1335.9090.02 ─ 09
13.4.3 UART trigger
Before you set up the trigger, make sure that the bus is configured correctly. See
Chapter 13.4.2, "UART configuration", on page 278.
To trigger on UART signals:
1. Press the [Protocol] key in the Analysis area of the front panel.
2. Select the bus that is configured for UART.
3. Select "Trigger".
This selection has several effects:
Enables decoding, if necessary.
Sets the "Trigger Type" to "Serial Bus" and the trigger source to the selected
bus.
Displays the "UART Trigger" condition in the dialog box, below the protocol
setup.
4. In the menu, select the "Source".
5. At "UART Trigger", select the required trigger condition:
"Start Bit" or "Frame Start": next start bit or first start bit after idle time
"Frame Start"
"Frame Error"
"Symbol <n>": frame number in a data stream
"Break": a start bit not followed by a stop bit
"Parity Error"
"Pattern": serial pattern of 1, 2, 3 or 4 symbols at a defined position in the data
stream
"Any Symbol": pattern of data bits anywhere in a data stream
6. If "Pattern" or "Any Symbol" is selected, the UART trigger setup dialog expands to
define the serial pattern or the symbol.
UART / RS232 (option R&S RTM-K2)

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Rohde & Schwarz RTM3000 Series and is the answer not in the manual?

Rohde & Schwarz RTM3000 Series Specifications

General IconGeneral
BrandRohde & Schwarz
ModelRTM3000 Series
CategoryTest Equipment
LanguageEnglish

Related product manuals